IBA - Automation of tests of stopping distances and accuracy of a robotic patient positioner with a laser tracker
Purnode, Florent
Promoteur(s) : Geuzaine, Christophe
Date de soutenance : 24-jui-2021/25-jui-2021 • URL permanente : http://hdl.handle.net/2268.2/12466
Détails
Titre : | IBA - Automation of tests of stopping distances and accuracy of a robotic patient positioner with a laser tracker |
Titre traduit : | [fr] IBA - Automatisation de tests de distances d'arrêt et de précision d'un positionneur de patient robotisé avec un traqueur laser |
Auteur : | Purnode, Florent |
Date de soutenance : | 24-jui-2021/25-jui-2021 |
Promoteur(s) : | Geuzaine, Christophe |
Membre(s) du jury : | Doubliez, Paul-François
Redouté, Jean-Michel |
Langue : | Anglais |
Nombre de pages : | 98 |
Mots-clés : | [fr] laser tracker [fr] robotics [fr] IBA [fr] protontherapy [fr] automation [fr] stopping distances [fr] patient positioned [fr] C++ [fr] Python [fr] Spatial Analyzer [fr] tests |
Discipline(s) : | Ingénierie, informatique & technologie > Ingénierie électrique & électronique |
Organisme(s) subsidiant(s) : | IBA (Ion Beam Applications) |
Public cible : | Professionnels du domaine |
Institution(s) : | Université de Liège, Liège, Belgique |
Diplôme : | Master : ingénieur civil électricien, à finalité spécialisée en "electronic systems and devices" |
Faculté : | Mémoires de la Faculté des Sciences appliquées |
Résumé
[fr] As a medical institution, IBA is requested to comply with medical standards. To verify them, it repeatedly needs to acquire a large amount of data through tests. These tests aim at checking the performances of the stopping distances of a robotic patient positioner. However, performing the tests manually is particularly repetitive and time-consuming. IBA thus wanted to automate these tests. This automation is the subject of this master thesis.
A complete automatic test system was developed. It allows for quickly retrieving stopping distance measurements at very low cost compared to the manual case. The final system also appears to be reliable, functional and user-friendly. Its development can be divided in two main parts: hardware and software developments.
The hardware development consisted in developing an electrical cabinet aiming at safely and properly interfacing all the test components together. These components are then piloted with the software part.
Once the system functional, the reliability of the measurements was checked. The automatic results appear to be either similar or more accurate than the manual ones.
Fichier(s)
Document(s)
Description:
Taille: 10.58 MB
Format: Adobe PDF
Annexe(s)
Description:
Taille: 1.91 MB
Format: Adobe PDF
Citer ce mémoire
L'Université de Liège ne garantit pas la qualité scientifique de ces travaux d'étudiants ni l'exactitude de l'ensemble des informations qu'ils contiennent.