Master thesis : Head phantoms: Design and implementation of a physiological signal generator on a PCB for the validation of new polysomnography systems
Boniver, Julien
Promotor(s) :
Redouté, Jean-Michel
Date of defense : 29-Jun-2026/30-Jun-2026 • Permalink : http://hdl.handle.net/2268.2/26100
Details
| Title : | Master thesis : Head phantoms: Design and implementation of a physiological signal generator on a PCB for the validation of new polysomnography systems |
| Translated title : | [fr] Conception et réalisation d’un générateur de signaux physiologiques sur PCB pour la validation de nouveaux systèmes de polysomnographie |
| Author : | Boniver, Julien
|
| Date of defense : | 29-Jun-2026/30-Jun-2026 |
| Advisor(s) : | Redouté, Jean-Michel
|
| Committee's member(s) : | Vanderbemden, Philippe
Sacré, Pierre
|
| Language : | English |
| Number of pages : | 78 |
| Keywords : | [en] biosignal playback [en] waveform generator |
| Discipline(s) : | Engineering, computing & technology > Electrical & electronics engineering |
| Institution(s) : | Université de Liège, Liège, Belgique |
| Degree: | Master : ingénieur civil électricien, à finalité spécialisée en Neuromorphic Engineering |
| Faculty: | Master thesis of the Faculté des Sciences appliquées |
Abstract
[en] Polysomnography (PSG) is the gold standard for diagnosing sleep disorders and relies on
the simultaneous overnight acquisition of several physiological signals. Validating new PSG
acquisition systems before they reach patients requires a testbench capable of reproducing
realistic multimodal signals under controlled conditions; yet existing solutions either target
a single modality, lack configurable artifact injection, or remain inaccessible because of cost.
This work presents the design, implementation, and characterization of an accessible, multi
modal physiological signal generator on a printed circuit board for the validation of new PSG
systems. The signals of interest in the scope of this work are EEG, EOG, ECG and EMG.
The device replays real recordings from the Haaglanden Medisch Centrum sleep staging
database through eight channels: seven single-ended (3 EEG, 2 EOG, EMG) and one differ
ential (ECG). Cardiac and ocular artifacts are first extracted through a method-per-artifact
pipeline, allowing them to be re-injected at user-defined gains. The hardware combines a 12
bit eight-channel DAC, a digitally controlled two-stage resistive divider sized per modality,
and precision buffers. Three playback modes (raw replay, artifact mixing, and on-demand
artifact injection), each with optional sleep-stage selection, are controllable from a PC ter
minal without reflashing. Signals are stored on an SD card, and dry-electrode connection is
supported through ten on-board conductive pads. The complete device costs approximately
200 euros.
The generator presents an output range from 62 µV to 79 mV with a best-case noise floor near
10 µV, a faithful reproduction on the ECG differential channel across all configurations. On
the single-ended channels, inter-channel coupling through the shared analog reference node
limits faithful simultaneous multichannel replay below the millivolt range; the mechanism is
analyzed and mitigation strategies are proposed. The supply noise is shown to be rejected by
the differential output topology, making the generator effectively independent of the choice
of power source. Overall, the work demonstrates that accessible, multimodal, hardware-level
PSG validation with configurable artifact injection is achievable, and provides a documented
basis for future versions.
File(s)
Document(s)
Annexe(s)
Cite this master thesis
The University of Liège does not guarantee the scientific quality of these students' works or the accuracy of all the information they contain.

Master Thesis Online


All files (archive ZIP)