Co-alignment of detectors in dual-channel spectrometers by optical coherence tomography

Télécharger
Chavet, Quentin ULiège
Promoteur(s) : Loicq, Jerôme ULiège ; Borguet, Benoit
Date de soutenance : 1-jui-2019/2-aoû-2019 • URL permanente : http://hdl.handle.net/2268.2/6998
Détails
Titre : Co-alignment of detectors in dual-channel spectrometers by optical coherence tomography
Titre traduit : [fr] Co-alignement des detecteurs de spectromètres à double canal par tomographie en coherence optique
Auteur : Chavet, Quentin ULiège
Date de soutenance  : 1-jui-2019/2-aoû-2019
Promoteur(s) : Loicq, Jerôme ULiège
Borguet, Benoit 
Membre(s) du jury : Blain, Pascal ULiège
Grodent, Denis ULiège
Langue : Anglais
Nombre de pages : 56
Mots-clés : [en] Spectrometer
[en] Optical coherence tomography
[en] ELOIS
[en] Detectors alignement
[en] Focal plane assembly
[en] Experimental test
[fr] Spectromètre
[fr] Tomographie par coherence optique
[fr] Alignement de détecteurs
[fr] Plan focal
[fr] Test expérimental
Discipline(s) : Physique, chimie, mathématiques & sciences de la terre > Aérospatiale, astronomie & astrophysique
Ingénierie, informatique & technologie > Ingénierie aérospatiale
Centre(s) de recherche : AMOS (Advance Mechanical and Optical Systems)
Intitulé du projet de recherche : Development of the ELOIS instrument by the AMOS company for the European Space Agency
Institution(s) : Université de Liège, Liège, Belgique
Diplôme : Master en sciences spatiales, à finalité spécialisée
Faculté : Mémoires de la Faculté des Sciences

Résumé

[en] This work presents a method for detectors co-alignment in dual-channel spectrometer using optical coherence tomography. The method is proposed for the ELOIS (Enhanced Light Offner Imaging Spectrometer) developed by the AMOS company (Advance Mechanical and Optical Systems). This company designs and builds optical components and instruments related to the space industry and one of their current project is the ELOIS instrument.

A crucial aspect in the development of the instrument is the correct alignment of the focal plane assembly components. During this phase, the two detectors of the spectrometer need to be correctly co-aligned with each other in the six degrees of freedom. This work presents a method using optical coherence tomography for the co-alignment of the detectors in tip, tilt and focus.

The first part explains the theoretical aspects of optical coherence tomography and how it can be applied to the ELOIS instruments.
The second part is the application of optical coherence tomography on a test bench replicating the components of the ELOIS. The procedure is detailed for this experimental test and an explanation of the results demonstrates the accuracy achievable with this method.
The third part of this work explains the differences between the experimental test and the ELOIS instrument. The procedure of co-alignment and the achievable accuracy are detailed in the case of this specific dual-channel spectrometer.


Fichier(s)

Document(s)

File
Access MasterThesis_Quentin_Chavet.pdf
Description:
Taille: 41.73 MB
Format: Adobe PDF

Auteur

  • Chavet, Quentin ULiège Université de Liège > Master sc. spatiales, à fin.

Promoteur(s)

Membre(s) du jury

  • Blain, Pascal ULiège Université de Liège - ULiège > CSL (Centre Spatial de Liège)
    ORBi Voir ses publications sur ORBi
  • Grodent, Denis ULiège Université de Liège - ULiège > Département d'astrophys., géophysique et océanographie (AGO) > Labo de physique atmosphérique et planétaire (LPAP)
    ORBi Voir ses publications sur ORBi
  • Nombre total de vues 96
  • Nombre total de téléchargements 1










Tous les documents disponibles sur MatheO sont protégés par le droit d'auteur et soumis aux règles habituelles de bon usage.
L'Université de Liège ne garantit pas la qualité scientifique de ces travaux d'étudiants ni l'exactitude de l'ensemble des informations qu'ils contiennent.